Patients with asthma need to be able to use their inhaler effectively

Professor Henry Chrystyn highlights the importance of checking inhaler technique, and the common difficulties patients with asthma encounter when using these devices


It has been shown that good inhaler technique is associated with better control of asthma,1–3 reduced use of inhaled corticosteroids, and significantly fewer acute exacerbations and hospital admissions.4 However, patients who have been prescribed inhalers may have problems with treatment compliance5-8 and performing the correct administration technique.9 National guidance therefore recommends that compliance and inhaler technique are checked prior to altering either dosage or inhaler device, and that instructions on correct administration of therapy should be provided regularly by a highly trained healthcare professional.10,11 Treatment compliance can be improved by patient education12 and satisfaction with the device,13 while training helps with inhaler technique;1,14 however many patients do not retain the trained method so regular inhaler technique checks are required.15

Background

The problems that patients have with the use of pressurised metered dose inhalers (pMDIs) were identified almost 50 years ago16-18 and still persist today.3 Dry powder inhalers (DPIs) were introduced to overcome these problems, but difficulties with administration of therapy have also arisen with these devices.19 The problems that patients with asthma encounter when using their inhalers have been emphasised by the Aerosol Drug Management Improvement Team (ADMIT) group20,21 and the International Primary Care Respiratory Group.22 A joint European Respiratory Society and International Society for Aerosols in Medicines (ERS/ISAM) Task Force was formed to raise awareness of these issues and highlight the inhalation techniques that patients should be trained to use.9 This article presents common problems that arise when patients use inhaler devices and advises on the techniques that all individuals should be trained to use.9

Pressurised metered dose inhalers

It has been shown that only 8% of adults with asthma are able to use their pMDI correctly, with a slow inhalation flow and good coordination.1 Both subjective assessments and objective measurements4,6,17,18,23 have highlighted errors that can occur when people use pMDIs, including ‘failure to coordinate actuation and inhalation’ (27% of users).24 Studies have also shown that reduced lung deposition of inhaled therapy can occur as a result of poor coordination between inhalation and actuation:25

  • when people do not exhale before an inhalation26 (exhaling before an inhalation increases lung deposition; many patients do not exhale before they use their inhaler4)
  • when a fast inhalation flow is used27,28
  • when individuals do not hold their breath after inhalation.27

The generic method for using a pMDI and breathe-actuated pMDI, as recommended by the ERS/ISAM Task Force is described in Box 1.9 It defines the traditional instruction of a slow and deep inhalation by a period of time. It is recommended that the inhalation phase should take 4–5 seconds in an adult, and 2–3 seconds in a child.9 The patient should be encouraged to perform a relaxed inhalation, and told that the dose should be actuated soon after they begin their slow inhalation. A recent study has confirmed that training patients to prolong their inhalation phase increases inhalation time without affecting the inhalation volume and so the inhalation flows are reduced, which is beneficial.1,29

Some pMDI products emit particles that are extra-fine and it has been shown that lung deposition of such very small particles is not affected by the inhalation flow.28 Furthermore studies have shown that for these extra-fine particle products coordination is not critical30 and so an early or late actuation is acceptable. In theory, this suggests that products with a formulation of extra-fine particles are less affected by a poor inhalation technique, but the clinical significance of this has not been studied. Some beclometasone and ciclesonide pMDIs emit extra-fine particles.


Patients with asthma need to be able to use their inhaler effectively continued

Box 1: Detailed instructions on how to use pMDIs, BA-pMDIs, pMDIs with spacers, and DPIs9

pMDIs: for patients with good actuation–inhalation coordination
1) Shake four or five times if suspension formulation.
2) Take the cap off.
3) Prime the inhaler (refer to the patient information leaflet for specific instructions).
4) Exhale slowly, as far as comfortable (to empty the lungs).
5) Hold the inhaler in an upright position.
6) Immediately place the inhaler in the mouth between the teeth, with the tongue flat under the mouthpiece.
7) Ensure that the lips have formed a good seal with the mouthpiece.
8) Start to inhale slowly, through the mouth and at the same time press the canister to actuate a dose.
9) Maintain a slow and deep inhalation, through the mouth, until the lungs are full of air. This should take an adult 4–5 seconds.
10) At the end of the inhalation, take the inhaler out of the mouth and close the lips.
11) Continue to hold the breath for as long as possible, or up to 10 seconds before breathing out.
12) Breathe normally.
13) If another dose is required, repeat steps 4–12.

BA-pMDIs: for patients ≥6 years old
1–7) Same as above for pMDIs alone.
8) Start to inhale slowly, through the mouth. The patient should sense that a dose has been released, either by taste or a noise when the dose is released (the noise is quiet for the Easi-Breathe®).
9) Maintain a slow and deep inhalation, through the mouth, until the lungs are full of air. This should take a child
2–3 seconds, and an adult 4–5 seconds.
10–13) Same as above for pMDIs alone.

pMDI + spacer with mouthpiece: for patients ≥6 years old (caregiver should determine if child can perform this technique correctly)
1) Shake four or five times if suspension formulation.
2) Take the cap off.
3) Prime the inhaler (refer to the patient information leaflet for specific instructions).
4) Insert the mouthpiece of the pMDI into the open end of the spacer and ensure a tight fit. If a reverse flow spacer is used,
insert the valve stem of the pMDI into the port on the mouthpiece of the spacer.
5) Place the mouthpiece of the spacer in the patient’s mouth with the teeth over the mouthpiece and the lips sealed around it.
6) Instruct the individual to exhale slowly, as far as comfortable (to empty their lungs).
7) Actuate one dose into the chamber of the spacer and start to inhale slowly through the mouthpiece. Some spacers will make a whistling noise if inspiration is too fast.
8) Maintain a slow and deep inhalation through the mouth, until the lungs are full of air. This should take a child 2–3
seconds, and an adult 5 seconds.
9) At the end of the inhalation, take the inhaler out of the mouth and close the lips.
10) Continue to hold the breath for as long as possible for up to 10 seconds before breathing out.
11) Breathe normally.
12) If another dose is required, repeat steps 1–11.
13) If inhaled corticosteroids are used, rinse mouth afterwards.

DPIs: for patients ≥5 years old (caregiver should determine if child can perform this technique correctly)
1) Take the cap off (some do not have a cap).
2) Follow the dose preparation instructions in the patient information leaflet.
3) Do not point the mouthpiece downwards once a dose has been prepared for inhalation because the dose could fall out.
4) Exhale slowly, as far as comfortable (to empty the lungs). Do not exhale into the DPI.
5) Start to inhale forcefully through the mouth from the very beginning. Do not gradually build up the speed of inhalation.
6) Continue inhaling until the lungs are full.
7) At the end of the inhalation take the inhaler out of the mouth and close the lips. Continue to hold the breath for as long
as possible, or up to 10 seconds.
8) Breathe normally.
9) If another dose is required, repeat steps 1–8.

pMDI=pressurised metered dose inhaler; BA-pMDI=breath-actuated pressurised metered dose inhaler; DPI=dry powder inhaler

Adapted from the European Respiratory Society. Eur Respir J June 2011; 37: 1308–1417; published ahead of print February 10, 2011, doi:10.1183/09031936.00166410. Reproduced with permission


Patients with asthma need to be able to use their inhaler effectively continued

pMDI plus spacers

Spacers were introduced to inhalers to help overcome problems with coordination. A cloud of aerosol is discharged from the pMDI into the spacer, and is then inhaled. After the pMDI is emitted into the spacer the aerosol cloud is stationary and so inhaling from this compared with inhaling a high-velocity spray discharged from a pMDI improves lung deposition,31,32 with greater penetration into the peripheral zones.33 This reduces the impact of the drug in the oro-pharyngeal area and is therefore beneficial. The ERS/ISAM Task Force recommendations on how to use a spacer are shown in Box 1.9

Most spacers are holding chambers and so can be used with tidal (relaxed normal) breathing. Only one dose should be discharged into the spacer—the use of multiple puffs results in a decrease in drug delivery.34 Only 3–5 relaxed breaths should be used when using tidal breathing because the particles will deposit as a result of gravity. If a child is using tidal breathing they should be encouraged to use one full inhalation as soon as they are able to perform this manoeuvre.35

As some spacers are prone to static, they should be washed in water that contains a detergent, and allowed to air dry to reduce the static effect.36 They should not be rinsed with water although instructions for some devices may specify rinsing in this way. It should be noted that spacers are the least preferred portable inhalation method because of their bulk.37

Dry powder inhalers

In recent years, DPIs have been more widely prescribed and used, and many different devices have been introduced. Each type of DPI has its own unique instructions for preparing a dose. Dry powder inhalers have different resistances and so the flows achieved should not be compared; for further details, see Box 2.

Incorrect dose preparation is more than likely to result in no dose being inhaled, irrespective of the inhalation manoeuvre performed by the patient. The ERS/ISAM Task Force has, therefore, emphasised the importance of following the recommended dose preparation instructions provided in the patient information leaflet for each device.9 Once a dose has been prepared, the mouthpiece of a DPI should not be tilted downwards because the dose, which is ready for inhalation, is likely to fall out. Other common errors made by patients using DPIs include:24

  • holding the device in the wrong position or incorrectly
  • actuation at total lung capacity/not exhaling to residual volume before inhaling
  • inadequate or no breath holding after inhalation
  • not using a fast inhalation from the start.

Patients should be trained to perform an inhalation as fast as they can and continue it for as long as possible, as indicated in the ERS/ISAM Task Force instructions in Box 1.9 This will enable them to receive the required lung dose during their inhalation, irrespective of the internal flow resistance of the device. Such inhalations will result in flows that are effective even in devices with a high resistance.47,48

The ERS/ISAM Task Force therefore stresses that the fast inhalation when using a DPI should commence immediately (i.e. at the start of each inhalation). It is important that patients who use an inhalation manoeuvre which starts slowly and gradually increases towards a peak rate should be trained to use a forceful inhalation from the start.

Healthcare professionals should exercise care when prescribing a DPI to young children, and the elderly with chronic obstructive pulmonary disease, as these two groups have reduced inspiratory capacity.38,49


Patients with asthma need to be able to use their inhaler effectively continued

Box 2: Inhalation flow in DPIs

After dose metering (during the dose preparation stage required before each inhalation), the drug particles in a DPI become deaggregated as a result of the turbulent energy generated by the interaction of the patient’s inhalation flow with the resistance inside the inhaler.38 The deaggregated dose leaves the DPI in the first part of an inhalation and the deaggregation process is related to the initial acceleration of the inhalation flow.24,39,40 Different DPIs do not have the same internal resistance to airflow and range from low to high resistance, and each DPI has a threshold below which deaggregation is not efficient.9

When comparing different devices it should be stressed that the inhalation flow only contributes to deaggregation, and that each type of DPI has a different resistance to achieve a set turbulent energy.9 If a DPI has a high resistance then a lower inhalation flow is required compared with a DPI that has a lower resistance. Additionally, higher internal energies are generated when patients use DPIs with a higher resistance because the relationship between the turbulent energy with flow and resistance is not linear.41

The common practice of comparing peak inhalation flow between different devices and suggesting that a DPI is not suitable for a patient because of low inhalation flow is therefore incorrect.38 However, it is valid to compare inhalation flows when using the same device: the higher the inhalation flow, the greater the turbulent energy, thereby leading to better deaggregation of the formulation. All DPIs demonstrate flow-dependent dose emission42 with better lung deposition at higher flows.43,44 It has been suggested that this relates to an improved clinical response, but this is not conclusive.44-46

DPI=dry powder inhaler

Inhalers in real-life settings

The importance of correct inhaler technique, as discussed in this article and highlighted by the ERS/ISAM Task Force, has been reinforced by a recent working party guideline on the primary care management of asthma in real-life settings, which was supported by Teva UK Limited through an educational grant.50 This guideline highlighted that using different devices for controller and reliever therapy may affect control of asthma and that it is therefore good practice for clinicians to prescribe the same inhaler type for both where possible.

When prescribing an inhaler for asthma, healthcare professionals should:50

  • be able to demonstrate appropriate use of inhaler devices to enable them to make a clinically informed decision
  • be aware that cost comparable devices are available, which patients may find easier to use
  • take into account patient factors, such as physical and mental disability, age, and reducing dexterity with advancing age.

The working party group recommended choosing an inhaler that the patient can and will use effectively, and that individuals are involved in the decision-making process. Commissioners of care should ensure that healthcare professionals receive appropriate training in the use of inhaler devices.50


Patients with asthma need to be able to use their inhaler effectively continued

GP commissioning take home messages for England

written by Dr David Jenner, NHS Alliance GMS contract/PBC Lead

  • Correct inhaler use is vital for the proper control of asthma, yet most adults do not use pMDIs effectively
  • The QOF incentivises practices to check and record inhaler use at least annually for patients with asthma or chronic obstructive pulmonary disease
  • CCGs could look to reduce avoidable emergency admissions and improve patient care by offering training for practice nurses as, with the abolition of PCTs, little primary care support is available from NHS England (formerly the NHS Commissioning Board)
  • Although inhaler devices vary in price, NICE guidance says that patients should be offered an inhaler of their own choice, which, if used correctly, is likely to be cost effective and reduce the likelihood of exacerbations
  • Local formularies for inhaler devices with cost comparisons, summaries of licensed indications, and dosing regimens are likely to help improve prescribing and help patients receive appropriate and effective treatments.

pMDIs=pressurised metered dose inhalers; QOF=quality and outcomes framework;
CCG=clinical commissioning group; PCTs=primary care trusts

Key points
  • Good inhaler technique is associated with improved control of asthma
  • Treatment compliance and inhaler technique should be checked before changing dosage or device
  • Patients can have difficulties using both pMDIs and DPIs
  • The European Respiratory Society and International Society for Aerosols in Medicines Task Force have detailed instructions on using inhaler devices
  • Patients should follow the instructions provided in the patient information leaflet for each device
  • Irrespective of inhaler type, patients should:
    • inhale gently as far as comfortable before their inhalation manouvre
    • use a relaxed inhalation
    • hold their breath after each inhalation
  • pMDIs:
    • the inhaler should be shaken before use
    • a slow inhalation (defined as an inhalation lasting for up to 5 seconds or longer in an adult and 2–3 in a child) should be performed
    • encourage patients to actuate a dose soon after the start of an inhalation
  • DPIs:
    • follow the dose preparation instructions in the patient information leaflet
    • the inhaler should not be pointed downwards after dose preparation
    • the patient should inhale as fast as possible from the start of an inhalation and continue for as long as possible
  • Healthcare professionals should prescribe an inhaler that patients can and will use.

pMDI=pressurised metered dose inhaler; DPI=dry powder inhaler


Patients with asthma need to be able to use their inhaler effectively continued

Conclusion

Many patients have problems using the correct inhaler technique. It is therefore important that they are prescribed an inhaler that they can and will use. Patients should comply with their inhaled medication and use the correct method of administration. Each type of inhaler has unique instructions for its use, and patients should be trained accordingly and their technique checked at regular intervals.

Pharmacists should be encouraged to perform medicines reviews and use the New Medicine Service as an opportunity to train patients to adopt the correct inhaler technique, and to check that they have done so.

If a patient is unable to demonstrate good inhaler technique with one device, an alternative inhaler that they can use should be identified.

References

  1. Al-Showair R, Pearson S, Chrystyn H. The potential of a 2tone trainer to help patients use their metered-dose inhalers. Chest 2007; 131 (6): 1776–1782.
  2. Giraud V, Roche N. Misuse of corticosteroid metered-dose inhaler is associated with decreased asthma stability. Eur Respir J 2002; 19 (2): 246–251.
  3. Kamps A, Brand P, Kimpen J et al. Outpatient management of childhood asthma by paediatrician or asthma nurse: randomised controlled study with one year follow up. Thorax 2003; 58 (11): 968–973.
  4. Melani A, Bonavia M, Cilenti V et al. Inhaler mishandling remains common in real life and is associated with reduced disease control. Respir Med 2011; 105 (6): 930–938.
  5. Rand C, Nides M, Cowles et al. Long-term metered-dose inhaler adherence in a clinical trial. The Lung Health Study Research Group. Am J Respir Crit Care Med 1995; 152 (2): 580–588.
  6. Cochrane M, Bala M, Downs K et al. Inhaled corticosteroids for asthma therapy: patient compliance, devices, and inhalation technique. Chest 2000; 117 (2): 542–550.
  7. Hoskins G, McCowan C, Neville R et al. Risk factors and costs associated with an asthma attack. Thorax 2000; 55 (1): 19–24.
  8. Restrepo R, Alvarez M, Wittnebel L et al. Medication adherence issues in patients treated for COPD. Int J Chron Obstruct Pulmon Dis 2008; 3 (3): 371–384.
  9. Laube B, Janssens H, de Jongh F et al; European Respiratory Society; International Society for Aerosols in Medicine. What the pulmonary specialist should know about the new inhalation therapies. Eur Respir J 2011; 37 (6): 1308–1331.
  10. British Thoracic Society, Scottish Intercollegiate Guidelines Network. British guideline on the management of asthma. SIGN 101. Edinburgh: SIGN, 2012. Available at: www.sign.ac.uk/guidelines/fulltext/101/index.html
  11. NICE website. Asthma quality standard. guidance.nice.org.uk/QS25 (accessed 3 April 2013).
  12. van der Palen J, Klein J, Rovers M. Compliance with inhaled medication and self-treatment guidelines following a self management programme in adult asthmatics. Eur Respir J 1997; 10 (3): 652–657.
  13. Small M, Anderson P, Vickers A et al. Importance of inhaler-device satisfaction in asthma treatment: real-world observations of physician-observed compliance and clinical/patient-reported outcomes. Adv Ther 2011; 28 (3): 202–212.
  14. Schulte M, Osseiran K, Betz R et al. Handling of and preferences for available dry powder inhaler systems by patients with asthma and COPD. J Aerosol Med Pulm Drug Deliv 2008; 21 (4): 321–328.
  15. Shim C, Williams M. The adequacy of inhalation of aerosol from canister nebulizers. Am J Med 1980; 69 (6): 891–894.
  16. Paterson I, Crompton G. Use of pressurised aerosols by asthmatic patients. Br Med J 1976; 1 (6001): 76–77.
  17. Oprehek J, Gayrard P, Grimaud C, Charpin J. Patient error in use of bronchodilator metered aerosols. Br Med J 1976; 1 (6001): 76.
  18. Saunders K. Misuse of inhaled bronchodilator agents. Br Med J 1965; 1 (5441): 1037–1038.
  19. Lavorini F, Magnan A, Dubus J et al. Effect of incorrect use of dry powder inhalers on management of patients with asthma and COPD. Respir Med 2008; 102 (4): 593–604.
  20. Crompton G, Barnes P, Broeders M et al; Aerosol Drug Management Improvement Team. The need to improve inhalation technique in Europe: a report from the Aerosol Drug Management Improvement Team. Respir Med 2006; 100 (9): 1479–1494.
  21. Broeders M, Vincken W, Corbetta L; ADMIT Working Group. The ADMIT series—Issues in Inhalation Therapy. 7. Ways to improve pharmacological management of COPD: the importance of inhaler choice and inhalation technique. Prim Care Respir J 2011; 20 (3): 338–343.
  22. Haughney J, Price D, Barnes N et al. Choosing inhaler devices for people with asthma: current knowledge and outstanding research needs. Respir Med 2010; 104 (9): 1237–1245.
  23. Goodman D, Israel E, Rosenberg M et al. The influence of age, diagnosis, and gender on proper use of metered-dose inhalers. Am J Respir Crit Care Med 1994; 150 (5 Pt 1): 1256–1261.
  24. Chrystyn H, Price D. Not all asthma inhalers are the same: factors to consider when prescribing an inhaler. Prim Care Respir J 2009; 18 (4): 243–249.
  25. Newman S, Weisz A, Talaee N et al. Improvement of drug delivery with a breath actuated pressurised aerosol for patients with poor inhaler technique. Thorax 1991; 46 (10): 712–716.
  26. Hindle M, Newton D, Chrystyn H. Investigations of an optimal inhaler technique with the use of urinary salbutamol excretion as a measure of relative bioavailability to the lung. Thorax 1993; 48 (6): 607–610.
  27. Newman S, Pavia D, Garland N, Clarke S. Effects of various inhalation modes on the deposition of radioactive pressurized aerosols. Eur J Respir Dis Suppl 1982; 119: 57–65.
  28. Usmani O, Biddiscombe M, Barnes P. Regional lung deposition and bronchodilator response as a function of beta-2 agonist particle size. Am J Respir Crit Care Med 2005; 172 (12): 1497–1504.
  29. Azouz W, Campbell J, Stephenson J et al. Improved metered dose inhaler technique when a co-ordination cap is used. J Aerosol in Med and Pulm Drug Deliv (in press).
  30. Leach C, Davidson P, Hasselquist B, Boudreau R. Influence of particle size and patient dosing technique on lung deposition of HFA-beclomethasone from a metered dose inhaler. J Aerosol Med 2005; 18 (4): 379–385.
  31. Newman S, Millar A, Lennard-Jones T et al. Improvement of pressurised aerosol deposition with Nebuhaler spacer device. Thorax 1984; 39 (12): 935–941.
  32. Newman S, Talaee N, Clarke S. Salbutamol aerosol delivery in man with the Rondo Spacer. Acta Ther 1991, 17: 49–50.
  33. Melchor R, Biddiscombe M, Mak V et al. Lung deposition patterns of directly labelled salbutamol in normal subjects and in patients with reversible airflow obstruction. Thorax 1993; 48 (5): 506–511.
  34. Clark D, Lipworth B. Effect of multiple actuations, delayed inhalation and antistatic treatment on the lung bioavailability of salbutamol via a spacer device. Thorax 1996; 51 (10): 981–984.
  35. Roller C, Zhang G, Troedson R et al. Spacer inhalation technique and deposition of extrafine aerosol in asthmatic children. Eur Respir J 2007; 29 (2): 299–306.
  36. Pierart F, Wildhaber J, Vrancken I et al. Washing plastic spacers in household detergent reduces electrostatic charge and greatly improves delivery. Eur Respir J 1999; 13 (3): 673–678.
  37. Lenney J, Innes J, Crompton G. Inappropriate inhaler use: assessment of use and patient preference of seven inhalation devices. EDICI. Respir Med 2000; 94 (5): 496–500.
  38. Azouz W, Chrystyn H. Clarifying the dilemmas about inhalation techniques for dry powder inhalers: integrating science with clinical practice. Prim Care Respir J 2012; 21 (2): 208–213.
  39. Everard M, Devadason S, Le Souef P. Flow early in the inspiratory manoeuvre affects the aerosol particle size distribution from a Turbuhaler. Respir Med 1997; 91 (10): 624–628.
  40. Kamin W, Genz T, Roeder S et al. Mass output and particle size distribution of glucocorticosteroids emitted from different inhalation devices depending on various inspiratory parameters. J Aerosol Med 2002; 15 (1): 65–73.
  41. Azouz W, Chetcuti P, Hosker H et al. Inhalation characteristics of patients when they use different dry powder inhalers (DPIs). J Aerosol in Med and Pulm Drug Deliv (in press).
  42. Chrystyn H. Is inhalation rate important for dry powder inhalers? Using the In-Check Dial to identify these rates. Respir Med 2003; 97 (2): 181–187.
  43. Borgstrom L, Bondesson E, Moren F et al. Lung deposition of budesonide inhaled via Turbuhaler: a comparison with terbutaline sulphate in normal subjects. Eur Respir J 1994; 7 (1): 69–73.
  44. Newman S, Moren F, Trofast E et al. Terbutaline sulfate Turbuhaler: effect of inhaled flow-rate on drug deposition and efficacy. Int J Pharm 1991; 74: 209–213.
  45. Pedersen S, Hansen O, Fuglsang G. Influence of inspiratory flow rate upon the effect of a Turbuhaler. Arch Dis Child 1990; 65 (3): 308–319.
  46. Nielsen K, Skov M, Klug B et al. Flow dependent effect of formoterol dry-powder inhaled from Aerolizer. Eur Respir J 1997; 10 (9): 2105–2109.
  47. Newhouse M, Nantel N, Chambers C et al. Clickhaler (a novel dry powder inhaler) provides similar bronchodilation to pressurized metered-dose inhaler, even at low flow rates. Chest 1999; 115 (4): 952–956.
  48. Koskela T, Malmstrom K, Sairanen U et al. Efficacy of salbutamol via Easyhaler unaffected by low inspiratory flow. Respir Med 2000; 94 (12): 1229–1233.
  49. Chrystyn H. Effects of device design on patient compliance: comparing the same drug in different devices. In: Dalby R, Byron P, Peart J et al. Respiratory Drug Delivery Europe 2009. River Grove: Davis Healthcare International Publishing, 2009: pp.105–116.
  50. Price D, Chrystyn H, Corrigan C et al. The primary care management of asthma in real-life settings. In: Hayeem N, editor.
    Guidelines—summarising clinical guidelines for primary care. 49th ed. Berkhamsted: MGP Ltd, February 2013. Available at: G

Professor Henry Chrystyn
Division of Pharmacy and Pharmaceutical Sciences
School of Applied Sciences
University of Huddersfield

View the Guidelines summary of the working party guideline on The primary care management of asthma in real-life settings at:

A checklist to aid the implementation of the working party guideline on Primary care management of asthma in real-life settings can be found on the downolads section of eGuidelines.co.uk or by clicking here.

The checklist covers:
  • inhaler technique
  • identification of patients with possible poor inhaler technique
  • low treatment adherence
  • young patients
  • asthma in smokers
  • managing allergic rhinitis in patients with asthma
  • other asthma phenotypes
  • stepping down treatment
  • specialist assessment.